Cytometry & Cell Sorting Laboratory Instruments & Services
Cytek Aurora Spectral Analyzer
Cytek Aurora Spectral Analyzer
The expert staff at NYU Langone Health’s Cytometry and Cell Sorting Laboratory can assist you in using the Cytek Aurora Spectral analyzer. Reserve this cell sorter using iLab (Kerberos ID and password required).
Cytek Aurora location
NYU Langone Health
550 First Avenue
Smilow Cellar, Room C32
Phone: 212-263-5907
Cytek Aurora Configuration
Cytek Aurora Biosafety Level
- BSL-1 (nonhuman or fixed human cells and cell lines)
Cytek Aurora Sample Preparation
- Use only 12 × 75 mm polystyrene sample tubes (polypropylene are not compatible).
- If using Automatic Micro-Sampling System (AMS): 96 well plates (U, V or flat bottom), 384 U bottom plates
- Concentration should be 1 × 106 cells/mL in at least 250 ul.
- Always bring extra media to dilute if necessary.
- All samples should be filtered
Cytek Aurora Pulse Processing
- Pulse area and height measurement available for all channels
- Pulse width of forward and side scatter
- Pulse width of one fluorescence parameter for each laser
Cytek Aurora Microparticle Detection
- Violet side scatter detector for thresholding on microparticles
Cytek Aurora Volumetric Measurement
- Direct volumetric measurement for cell counting